Analysis › MINFLUX · version 1
How the MINFLUX sequence ran: the attempts passed and lost per
Shows how a MINFLUX measurement ran: how many of the molecules the microscope started to localize it lost on the way and why, how much background it saw, how close its first guesses were to where each molecule turned out to be, how many localizations and photons each molecule gave, and where the time went. These are the plots abberior's ParaFlux draws, and the ones its wiki tells a MINFLUX user to look at before trusting a measurement or changing the sequence or the buffer.
MINFLUX localizes a molecule in steps, called iterations. The first, the scouting iteration, looks for a molecule with a wide beam pattern. If it finds one, the catching iterations close in on it, each with a smaller pattern. The final iteration, with the smallest pattern, then repeats for as long as the molecule stays on, and each repetition is one localization. An attempt can end at any iteration: too few photons (the molecule went dark), a failed centre-frequency-ratio check (cfr, a sign of two molecules on at once), or a position out of the scanner's range. The localizations of one molecule while it was on form a trace (tid), which ParaFlux calls a burst.
It needs the file opened with every iteration ticked (File > Open, or the Load plugin's Abberior MSR part): a normal table holds only the localizations, not the attempts that failed or the iterations before. It reads every row of the files the layer shows, whatever the layer's filter: that filter starts on the localizations (final = 1), and the attempts that failed have no position for a filter or a drawn ROI to keep.
1. The attempts, iteration by iteration. Every scouting event is an attempt. For each iteration before the final one, it counts how many attempts reached it and how they ended there, as a share of all the attempts: the 100 % line is every molecule the scouting iteration looked at. Why an iteration ended is the instrument's status code (sta): passed, failed the cfr check, dark (no signal) or out of range. The final iteration repeats, so it is counted per row instead: the share of its rows that were valid localizations is in the panel's title.

Simulated attempts with known pass rates (8 %, 90 % and 60 % in the three iterations before the final one, so 8 %, 7.2 % and 4.3 % of the attempts pass): the bars and the crosses, the truth, agree. Most attempts end in scouting, dark – in a real measurement too, where most of the scouting iteration's rows find nothing. The losses in iteration 2 are 80 % dark and 20 % failed cfr checks, as simulated.
2. The background. Imspector estimates the background (fbg) in each iteration where background sensing is on; the median and quartiles per iteration are drawn. With background sensing off – as in many files – there is nothing to draw, and the panel says so.
3. The catching offsets. Each catching iteration gives a position. How far it lies from the localization it led to – the first localization of the trace after it – shows how well the sequence closes in: the offsets should shrink with each iteration and scatter around zero. A mean far from zero in one direction points to a misaligned pattern or pinhole; a large spread in the first iterations is expected, since their patterns are wide.

The offset of each catching iteration's position from the localization it led to, mean ± standard deviation per axis, for attempts simulated to scatter by 60, 20 and 8 nm in iterations 0, 1 and 2 (grey bars), about molecules localized to 4 nm.
4. Bursts: localizations and photons. For every trace that gave at least one localization (a valid burst) it counts the localizations and adds up the photons of all its rows, the failed iterations included (eco + ecc, the photon budget). The number of localizations per burst is roughly exponentially distributed; its mean is the estimate drawn. ParaFlux's guidance: around 15 for immunostained cells, more for DNA-PAINT, under 10 means molecules are lost early.

Localizations per valid burst for bursts simulated with a mean of 15: the estimate comes back.
5. Where the time goes, and the parameters per iteration. Each row lasts until the next one starts; adding that up per iteration gives the share of the acquisition each iteration took – usually most of it in scouting, looking for molecules. Any attribute (per iteration, the photons by default) is drawn per iteration as median and quartiles, and along the first first localizations of every trace, which shows a trace's start settling in.
6. efo against cfr. For the localizations whose cfr was measured, their photon rate (efo) against their cfr: one molecule gives one cloud at low cfr; two molecules on at once give a second one at higher cfr and higher efo. In 2D sequences cfr is measured once a trace, so the panel is a scatter of few points.
The final iteration is the highest iteration that holds a localization (a row with final = 1: valid and marked final by the instrument). The iterations below it are counted per attempt: the share of the attempts that reached iteration k and ended with outcome o is the number of rows of iteration k with that outcome over the number of rows of iteration 0. In a 3D sequence, whose iterations alternate between xy and z, every iteration is counted; ParaFlux draws only the complete xy-and-z steps.
The outcomes come from sta, which the two record layouts code differently (abberior's wiki, "MINFLUX Data Parameters"): m2410 1 valid final, 2 valid, 6 failed cfr check, 8 no signal, 9 DAC out of range, 11 background measurement, 4 and 5 derived or reserved; m2205 0 valid, 4089 cfr, 4096 no signal, 4100 out of range. A code not in either list counts as passed where the row is valid, else as other.
Why a trace ended is its last row's code, a background measurement (11) aside. A valid burst whose last row is a localization ended with the measurement.
The catching offsets pair each non-final row that has a position with the localization of the same trace that has the same trace_index – the number of the localization a row leads to. The offset is the row's position minus that localization's; the mean and the population standard deviation are taken per iteration and axis.
Localizations per burst: the mean over the valid bursts, which is the maximum-likelihood estimate of a geometric distribution's mean.
The time per row is the time to the next row, all rows sorted by time_s; the last row gets none.
Per trace it also keeps a region table of tid, "per tid: Sequence QC": each trace's rows (n_rows), localizations (n_localizations), photon budget, duration and its last status (end_sta), for the region filter and the summaries.
| setting | default | what it does |
|---|---|---|
per iterationparameter | photons | The attribute shown per iteration and against the localization's number in its trace.
|
first localizationsfirst_localizations | 25 | How many of each trace's localizations the last panel follows. About 25 shows a trace's start creeping into place, which in 10-30 % of traces lasts 5 to 30 localizations; longer only averages fewer traces. 2 to 500 |
Nine panels: the attempts per iteration, the background per iteration, the catching offsets, localizations per burst, photons per burst, the chosen attribute per iteration, efo against cfr, the time per iteration, and the chosen attribute along each trace's start. The text gives the attempts, localizations and bursts, the share passed per iteration, why the bursts ended, and the scouting's share of the time. The numbers are kept with the result for a batch to compare measurements: passed_itr{k}, localizations_per_event, fbg_itr{k}_khz, locs_per_burst, photons_per_burst, ended_dark and the other reasons, scouting_time_fraction.
A good measurement of a sparse, well-blinking sample passes 40-50 % of the attempts that reach the catching iterations into the final one (ParaFlux's rule of thumb for the last bar), loses few to the cfr check, and gives bursts of 15 or more localizations. Large cfr losses mean too many molecules on at once (the buffer); large dark losses, too little excitation or molecules that blink off too soon.