← back to the tutorial

MINFLUX: opening, checking and analysing traces

The words of this tutorial, step by step. The text lives in src/smappy/tutorial/topics/; the directions it follows are in src/smappy/tutorial/STYLE.md.

MINFLUX in SMAPpy

1. card: MINFLUX: opening, checking and analysing traces

File, Open reads abberior's .msr: each measurement becomes a file, with its sequence.

Analysis, MINFLUX: what only MINFLUX data has, the sequence, its quality, two colours, the beads.

Analysis, Tracking: traces, from any method: their statistics, their steps, one trace up close.

Regions: the numbers per trace, compared between measurements.

The file here is simulated: motors stepping 8 nm, so the answers are known.

spoken: MINFLUX files open from the File menu. The tools are in three places: MINFLUX for what only MINFLUX has, Tracking for traces, and Regions to compare. The data is simulated, so the answers are known.

2. card: Words abberior uses

iteration: one step of the search: the beam pattern shrinks around the molecule, and the last iteration gives the localizations

sequence: the iterations and their settings, what the instrument was told

trace, tid: the localizations of one molecule while it was caught; tid numbers them

efo: the photon rate while localizing: twice as high with two molecules

cfr: how much the centre of the pattern still sees: high when a second molecule is near

thread: a sequence of its own in the same measurement, often a second colour

MBM: beads tracked during the measurement, for the drift

spoken: A few of abberior's words, which the plugins use too: iteration, sequence, trace, efo, cfr, thread and MBM.

Opening an .msr

3. Opening an .msr asks which parts to read. This simulated file holds two measurements of motors, one at high and one at low ATP.

4. Beside it, the selected measurement's sequence: what the instrument did in each iteration.

5. Every iteration reads every step of every attempt, not only the localizations. Sequence QC needs it; the layers still show only the localizations.

6. Each measurement is a file. With every iteration read, the layer keeps the localizations: final is 1.

7. A layer starts on the first file. Choose filenumber and tick all to see both measurements.

8. A tracking measurement opens as its tracks: every localization, each trace a line in its own colour. The motors walk.

The sequence

9. Sequence table is not on this tab at first: the + at its top finds any plugin and pins it here.

10. Sequence table lists each measurement's sequence, an iteration a line, and first what differs between them. Here nothing: the difference is in the sample.

11. File conditions writes that difference onto the localizations, here the ATP concentration read from the names.

How the sequence ran

12. Sequence QC reads every iteration: how many attempts passed each one, and why the others ended.

13. Most attempts end in the first iteration, where nothing is there yet. Once caught, a molecule mostly stays.

14. efo against cfr: one molecule is a cloud at low cfr; two molecules at once sit higher on both.

Quality

15. Quality filter finds the cuts in the data: efo below the valley before the two-molecule peak, and cfr below the single molecules' spread.

16. cfr is measured once, at the start of a trace, so it is cut per trace: a trace is kept or dropped whole.

17. The cuts are ordinary filter bounds on the layer: they can be moved or cleared like any other.

Traces

18. Trace statistics measures every trace at once: its speed, its shape, and its precision from consecutive localizations.

19. Each trace's numbers are kept as a table per tid, for filtering and comparing traces.

20. Trace inspector shows one trace up close: the longest, or the one you choose by its tid.

Steps

21. Find steps fits steps to each trace along its own axis. With the expected size set, steps that are too small are merged.

22. The longest trace, its positions along the track and the steps found in them.

23. The step sizes pooled over the traces, and the dwell times between steps, with three models compared.

Comparing the measurements

24. Compare groups compares what was measured per trace, grouped by file or by a condition: here ATP.

25. The steps are 8 nm in both. With little ATP the motor waits longer, and one event, ATP binding, sets the pace: the Erlang shape drops towards 1.

Next

26. card: Also there

- MINFLUX dual colour: two colours registered, from beads or the molecules, and their traces joined into molecules.
- MINFLUX kinesin (MINFLUX Workflow): from two-colour tracks to the steps of every walking kinesin, in one run.
- MBM drift correction and MBM registration: the beads, to correct drift and to put measurements on top of each other.
- Stability: how still the instrument was, from beads.
- Filter regions: keep the traces whose numbers pass a condition.Each plugin's ? opens its page.

spoken: There is more: two colours, the beads for drift, the instrument's stability, and filtering traces by their numbers. Each plugin's question mark explains it.